Continuous, intrinsic magnetic depletion of erythrocytes from whole blood with a quadrupole magnet and annular flow channel; pilot scale study.

Moore LR, Mizutani D, Tanaka T, Buck A, Yazer M, Zborowski M, Chalmers JJ
Biotechnol Bioeng 115 1521-1530 01/01/2018

Abstract

The ability to separate RBCs from the other components of whole blood has a number of useful clinical and research applications ranging from removing RBCs from typical clinical blood draw, bone marrow transplants to transfusions of these RBCs to patients after significant blood loss. Viewed from a mechanistic/process perspective, there are three routine methodologies to remove RBCs: 1) RBCs lysis, 2) separation of the RBCs from the nucleated cells (i.e., stem cells) based on density differences typically facilitated through centrifugation or sedimentation agents, and 3) antibody based separation in which a targeted RBC is bound with an affinity ligand that facilitates its removal. More recently, several microfluidic based techniques have also been reported. In this report, we describe the performance of continuous RBC separation achieved by the deflection of intrinsically magnetic, deoxygenated RBCs as they flow through a magnetic energy gradient created by quadrupole magnet. This quadrupole magnetic, with aperture of 9.65 mm, has a maximum field of B

Full Text