Comparing intraoperative parathyroid identification based on surgeon experience
versus near infrared autofluorescence detection - A surgeon-blinded multi-centric study.

Thomas G, Solórzano CC, Baregamian N, Mannoh EA, Gautam R, Irlmeier RT, Ye F, Nelson JA, Long SE, Gauger PG, Magner A, Metcalf T, Shirley LA, Phay JE, Mahadevan-Jansen A

BACKGROUND : Near infrared autofluorescence (NIRAF) detection has previously demonstrated significant potential for real-time parathyroid gland identification. However, the performance of a NIRAF detection device - PTeye® - remains to be evaluated relative to a surgeon's own ability to identify parathyroid glands.

METHODS : Patients eligible for thyroidectomy and/or parathyroidectomy were enrolled under 6 endocrine surgeons at 3 high-volume institutions. Participating surgeons were categorized based on years of experience. All surgeons were blinded to output of PTeye® when identifying tissues. The surgeon's performance for parathyroid discrimination was then compared with PTeye®. Histology served as gold standard for excised specimens, while expert surgeon's opinion was used to validate in-situ tissues.

RESULTS : PTeye® achieved 92.7% accuracy across 167 patients recruited. Junior surgeons (<5 years of experience) were found to have lower confidence in parathyroid identification and higher tissue misclassification rate per specimen when compared to PTeye® and senior surgeons (>10 years of experience).

CONCLUSIONS : NIRAF detection with PTeye® can be a valuable intraoperative adjunct technology to aid in parathyroid identification for surgeons.